Spittlebugs (superfamily Cercopoidea) live within a mass of frothy, spittle-like foam that is produced as a by-product of their xylem-feeding habits. The wet spittle represents a unique respiratory environment for an insect, potentially acting either as a reserve of trapped oxygen (O2) or as a significant barrier to O2 diffusion from the surrounding atmosphere. Feeding on xylem sap under tension is also assumed to be energetically expensive, potentially placing further constraints on their gas exchange. To understand the respiratory strategies used by spittlebugs, this study measured the PO2 within the spittle of the meadow spittlebug, Philaenus spumarius, as well as the non-feeding metabolic rate (RMR) and respiratory quotient (RQ) of both nymphs and adults. The metabolic rate of nymphs feeding on xylem was also measured. In separate experiments, the ability of a nymph to obtain O2 from bubbles while submerged in foam was determined using a glass microscope slide coated in an O2-sensitive fluorophore. We determined that P. spumarius breathes atmospheric O2 by extending the tip of its abdomen outside of its spittle, rather than respiring the O2 trapped in air bubbles within the foam. However, spittlebugs can temporarily use these air bubbles to breathe when forcibly submerged. O2 and CO2 did not differ statistically within life stages, giving a RQ of 0.92 for nymphs and 0.95 for adults. Feeding on xylem was found to increase the nymphs' CO2 by only 20% above their RMR. From this cost of feeding, cibarial pump pressures were estimated to be between −0.05 and −0.26 MPa.

Spittlebugs are the nymph stage of the froghopper (Cercopoidea: Homoptera) and are so called because of the conspicuous mass of wet, spittle-like foam that they secrete around themselves while feeding on the sap of plants. Spittlebugs produce this foam by incorporating bubbles of air into an anal exudate that consists primarily of excreted xylem fluid – a plentiful by-product of their xylem-feeding habits (Wiegert, 1964). This fluid is further modified by the addition of proteins and mucopolysaccharides secreted by the Malpighian tubules, which are believed to stabilise the structure of the foam (Marshall, 1966, 1973; Mello et al., 1987). Bubbles are produced when the spittlebug extends the tip of its abdomen through the surface of the foam and draws atmospheric air into a groove on the ventral surface of its abdomen (Hamilton, 1982). This sternal groove is formed by overlapping pleural prolongations, and contains the nymph's abdominal spiracles (Guilbeau, 1908; Hamilton, 1982). Then, by submerging the tip of the abdomen into the foamy liquid and contracting the abdomen (Fig. 1A,B), bubbles are expelled. By rapidly repeating this action, a spittlebug can completely cover its body in foam within 15–30 min (Hamilton, 1982).

Fig. 1.

The meadow spittlebug, Philaenus spumarius. Photographs showing the extensibility of the abdomen of a P. spumarius spittlebug (A, contracted; B, extended), and a bubble chamber inside a spittle mass (C). The foam covering the chamber has been removed to expose the newly moulted adult froghopper within.

Fig. 1.

The meadow spittlebug, Philaenus spumarius. Photographs showing the extensibility of the abdomen of a P. spumarius spittlebug (A, contracted; B, extended), and a bubble chamber inside a spittle mass (C). The foam covering the chamber has been removed to expose the newly moulted adult froghopper within.

The adaptive function of the spittle mass is still debated. It has been proposed that it could protect the nymph from desiccation and/or predation (del Campo et al., 2011; Whittaker, 1970), provide a microclimate that buffers temperature (Tonelli et al., 2018), prevent growth of fungi (Li et al., 2015), or even act as a neutraliser of toxic ammonia (Rakitov, 2002). But, regardless of the potential adaptive benefits, living submerged in a mixture of liquid and gas is an unusual respiratory environment for any terrestrial insect. However, no study has yet determined precisely how spittlebugs obtain O2 and excrete CO2 while submerged in a wet foam (Wiegert, 1964). There are two main gas exchange strategies open to a spittle bug: either they respire the O2 trapped within the bubbles that comprise the foam, thus making the foam mass analogous to the air bubbles and compressible gas-gills carried by some diving insects (Turner, 1994), or they rely on a respiratory siphon or snorkel to access atmospheric O2 as required (Hamilton, 1982; Kosztarab, 1979). The primary goal of this study was to determine which of these strategies are used. Are they ‘foam breathers’ or do they snorkel?

There is one life stage of the spittlebug, however, where they must rely on the diffusion of O2 through the wet foam to breathe. Before their final moult from nymph to adult, many spittlebug species create a single large bubble within the foam mass (Fig. 1C). Within this ‘bubble chamber’, they moult into the adult froghopper, remaining concealed until their cuticle has hardened (Weaver and King, 1954). Unlike the spittlebug nymph, the adult froghopper has neither a flexible abdomen able to telescope out of the spittle nor the sternal groove, and thus it is unable to access atmospheric O2 directly while within the bubble chamber.

The mode of gas exchange used by spittlebugs is also of interest when considering their energetic demands. Spittlebugs feed by inserting their stylets into a plant's xylem tissue then drawing up the sap. However, the cohesion–tension theory of sap transport assumes that xylem sap can exist under very high tensions (>1 MPa) (Kim, 2013). If this is the case, then spittlebugs must expend a large amount of energy to produce an even greater negative pressure to draw out this fluid (Kim, 2013; Malone et al., 1999). All feeding suction must be generated by the bug's cibarial pump, which is powered by large muscles in their head (Malone et al., 1999). In addition, xylem sap is very poor in nutrients, so spittlebugs must ingest large quantities to meet their metabolic demands, routinely excreting 150–280 times their own body mass in fluid every 24 h (Horsfield, 1978). As the power (P; W) required to pump a fluid is equal to the volumetric flow rate of the fluid through the pump (Q; m3 s−1) and the pressure (p; N m−2) it works against (i.e. P=p×Q), xylem-feeding bugs would appear to be uniquely disadvantaged in their choice of food. Thus, it must be energetically very costly for spittlebugs to extract these large volumes of xylem sap, unless they employ some unknown method to reduce the xylem tension (Kim, 2013; Malone et al., 1999; Novotny and Wilson, 1997). Furthermore, if the increased metabolic requirements of feeding dramatically increased O2 consumption, then using the spittle mass as its primary oxygen source may limit the bug's ability to feed. However, the metabolic rate of a spittlebug feeding on xylem has never been measured, so the metabolic cost of this activity, and its consequences for gas exchange, remain unknown.

This study examined the gas exchange strategies and metabolic rates of both the nymph and adult of the meadow spittlebug Philaenus spumarius (Linnaeus 1758). Fibre optic O2 optodes were used to determine whether a decrease in PO2 was detectable within spittle masses containing a nymph, as would be consistent with the spittlebug respiring O2 from within the trapped air bubbles. This technique was also used to measure the PO2 inside the bubble chamber of a moulting final instar. Both stop-flow O2 respirometry and flow-through CO2 respirometry were used to determine the resting metabolic rate (RMR; CO2, O2) and respiratory quotient (RQ) of both the spittlebugs and adult froghoppers. Flow-through CO2 respirometry was also used to measure the metabolic rate of spittlebugs during locomotion and while producing spittle and feeding on xylem sap. Once the spittlebugs had constructed a spittle mass within the respirometry chamber, their responses to mechanical taps simulating the presence of a potential predator were also measured. Finally, spittlebugs were forcibly submerged in spittle on a microscope slide coated with an O2-sensitive fluorophore to test their ability to access O2 within the foam, and visualise how the PO2 changed within adjacent bubbles in the foam mass. Altogether, these measurements provide a comprehensive assessment of the energetic costs and gas exchange strategies that spittlebugs employ to live within the unusual respiratory environment of a wet foam.

Insect collection and identification

All spittlebug nymphs and adults used in these experiments were collected by hand from around the University of British Columbia Point Grey campus between May and July in 2017 and 2018. Individuals were identified using the key in Hamilton (1982), and the instars were determined by wing pad development using the key in Weaver and King (1954).

Field measurements of PO2 within established masses of spittlebug foam

PO2 transects through masses of P. spumaris spittlebug foam were measured at the Point Grey Campus of the University of British Columbia. Spittle masses of varying size and on various plant species were selected haphazardly. The PO2 within the spittle mass was determined using a needle-mounted O2 optode (130 μm diameter fibre, NTH-PST7-O2-L2.5-TS-NS40-0.8-NOP, PreSens GmbH, Regensburg, Germany) held in a 3-axis micromanipulator (M3301R, World Precision Instruments, Sarasota, FL, USA), which was mounted on a metal stake. The stake was driven into the ground next to the plant, and a second clamp attached to the stake was used to hold the plant's stem. The tip of the optode was always inserted into the spittle mass parallel to the stem of the plant, to avoid breaking the sensor. The optode was advanced into the spittle mass in 1 mm increments, from the outer edge of the mass to the centre where the nymph(s) resided, while PO2 measurements were recorded at each position using a portable fibre optic oxygen meter (Microx4, PreSens GmbH).

Laboratory measurements of PO2 within newly secreted foam

Spittlebugs collected in the field were quickly returned to the laboratory, where they were placed on potted sheep sorrel plants (Rumex acetosella) growing in a standard potting compost (Indoor Potting Soil, GardenWorks, BC, Canada) in 6 inch diameter by 4.5 inch deep pots. Spittlebugs were observed feeding on sheep sorrel in the field, and would readily feed and secrete foam masses on the potted plants. Once foam began to be produced, a needle-mounted O2 optode held by a micromanipulator was again inserted into the freshly produced foam, this time held perpendicular to the plant stem or leaf, with the tip positioned as close as possible to the feeding bug. The optode remained in this position for 2 h while PO2 was logged every 20 s.

Bubble chamber PO2

Plant stems and leaves that had bubble chambers containing a moulting final instar nymph on them were removed and taken back to the lab where the PO2 within the bubble chamber was recorded, again using a needle-mounted O2 optode. Each bubble chamber was inspected to make sure it was intact and completely covered by a layer of foam. The tip of the O2 optode was then carefully inserted into the bubble so as to not cause it to collapse or burst, and the PO2 within the bubble chamber was measured every 20 s for 2 h. At the end of the measurement period, the adult insect was removed from within the bubble, and its weight and length were measured. The reported PO2 values inside the bubble chambers were averaged from the first 15 min of measurements and were compared with the atmospheric O2 directly outside the bubble chamber.

Respirometry

The rates of O2 consumption (O2) and CO2 production (CO2) of spittlebug nymphs and adults at rest were determined in the laboratory. All nymphs were individually weighed to 0.01 mg on an electronic balance (XPE205 DeltaRange, Mettler Toledo, Greifensee, Switzerland) before and after respirometry measurements, and all adults were weighed after. Nymphs or adults were placed individually into 2 ml glass vial chambers (11 mm diameter, 35 mm long) which were then sealed and left for 2 h while the PO2 or CO2 concentration within the vial was continuously measured. A small piece of nylon mesh was also placed inside the vial to give the insects a surface to grip onto during the trials. During the measurement period, both the CO2 and O2 respirometry chambers were inserted into close-fitting holes bored horizontally into an aluminium block (45 mm×45 mm×146 mm). The block was placed on a temperature-controlled Peltier plate (AHP-1200CPV, ThermoElectric Cooling America Corporation, Chicago, IL, USA) and maintained at a constant temperature of 20°C.

The resting O2 of spittlebugs was determined using constant volume respirometry. To do this, an O2 sensor spot (SP-PSt7-10-NAU-D5-YOP, PreSens GmbH) was affixed to the bottom of a 2 ml glass vial and measured using a flat-broken optic fibre (POF-L2.5-1ST, PreSens GmbH), which was held against the outside of the vial while connected to a Microx4 O2 meter. A small hole through the bottom of the vial cavity in the aluminium block allowed the optic fibre to be inserted up against the bottom of the glass vial, and held directly opposite the sensor spot. The sensor spot was calibrated every day by continuously flushing the vial with compressed N2 gas (for the zero), then humidified, CO2-free air (for 100% air saturation) produced by a regenerative purge gas generator (CDA4, Puregas, Broomfield, CO, USA). Briefly, the dry, CO2-free air produced by the purge gas generator was first scrubbed of any remaining trace of CO2 or water vapour by passing it through 1 l columns of soda lime and Drierite (W. A. Hammond Drierite Co. Ltd, Xenia, OH, USA). Then, a mass-flow controller (Model MC-100SCCM-D/5M, Alicat Scientific, Tucson, AZ, USA) metered this air at 30 ml min−1 through an 18 cm length of Nafion tubing (TT-070 Nafion tubing, CD Nova, Surrey, BC, Canada) submerged in a 100 ml container containing reverse osmosis (RO) water, humidifying it to saturation at 20°C. Before the respirometry began, the insect was added to the vial, which was then flushed with the humidified CO2-free air at 30 ml min−1 for at least 1 min before the lid was screwed on to seal the vial. The percentage O2 in the vial was recorded every 30 s for the entire 2 h experiment. To calculate O2, first, the percentage O2 was converted to a volume of O2 (VO2, ml) by multiplying by the total volume of the vial (minus the volume of the nylon mesh and nymph inside) using the equation:
formula
(1)
This VO2 was then converted into VO2 STPD (standard temperature and pressure dry):
formula
(2)
where Pb is the ambient barometric pressure (kPa), Pwv is the saturated water vapour pressure at 20°C (kPa) and Ta is the ambient temperature (°C). The VO2 STPD within the vial was plotted over time and the slope of this change was the calculated O2, which was then converted to µl h−1.
The resting CO2 of adults and nymphs was determined with a flow-through respirometry system using a 2 ml vial. Two 21-gauge stainless steel hypodermic needles with Luer fittings were epoxied into holes in the vial's lid, as the inlet and outlet port. To prevent the spittlebugs from drying out over the 2 h experiment, the dry, CO2-free incurrent air stream was humidified as described above, by passing it through a length of Nafion tubing submerged in water, before being flushed through the chamber at 30 ml min−1. The air leaving the respirometry chamber was dried by passing it through a second length of Nafion tube in a shell-and-tube configuration; the Nafion tube was mounted within a larger-diameter acrylic tube that was continuously flushed in a countercurrent direction with 100 ml min−1 of dry air. This removed all water vapour from the excurrent airstream before it entered a CO2/H2O infra-red gas analyser (LI-7000, Licor, Lincoln, NE, USA). The LI-7000's auxiliary voltage output (configured for μmol CO2 mol−1, equivalent to concentration in ppm) was sampled at 2 Hz by Powerlab 8/35 DAQ analog to digital converter (ADInstruments, Bella Vista, NSW, Australia) which was recorded by LabChart (v.8.1.5, ADInstruments) running on a PC. CO2 was then calculated by taking the average CO2 ppm values over the last 1 h 45 min of the 2 h experiments. These averages were then converted to CO2 in µl h−1 using the equation:
formula
(3)
where i is incurrent flow rate (μl h−1) and FeCO2 is the excurrent CO2 concentration (ppm).

A similar flow-through respirometry system to that described above was used to measure CO2 during active xylem feeding and spittle production, where a clear acrylic chamber was used instead of the glass vials. The chamber was made from two square (50×50 mm) acrylic blocks with circular holes milled into each half (30 mm diameter and 5 mm deep in one half, 30 mm diameter and 6 mm in the other half, giving an approximate volume of 7.76 ml). The two squares interlocked to make a central circular chamber. A groove (3×3×50 mm) was cut across the midline of each acrylic block, bisecting the circular chamber. This groove allowed both blocks to be clamped around the stem of a plant, forming an enclosed chamber. The two halves of the chamber were sealed around a stem of sheep sorrel using 2-part polyvinylsiloxane casting material (President light body dental impression material, Coltène Whaledent, Altstätten, Switzerland). One half of the chamber had a round glass window epoxied into it for observation using a digital camera (PiNoir IR camera Raspberry Pi 3rd Edition Model B, Allied Electronics Inc., London, UK). The other half had two short lengths of 21-gauge stainless steel hypodermic tubing (8 mm long) epoxied into two holes in the top of the chamber, for the air inlet and outlet. Dry, CO2-free air was metered at a rate of 50 ml min−1 using a 0–100 ml min−1 mass flow controller (Alicat Scientific). Air was humidified before, and dried after, the respirometry chamber using the same Nafion tubes as described above. The dried excurrent air was passed into a LI-820 CO2 IRGA gas analyser (Licor) and CO2 concentration (in ppm), along with chamber temperature measured using a 0.010 gauge T-type thermocouple, were sampled at 2 Hz using a Powerlab 8/35 DAQ (ADInstruments) and recorded using LabChart software (v.8.1.5, ADInstruments).

After recording a CO2 baseline with a sheep sorrel stem mounted in the chamber, a spittlebug was introduced into the chamber through a port (4 mm diameter) in the acrylic block. This opening was then sealed with putty (UHU GmbH & Co. KG, Bühl, Germany). Spittlebug behaviour and movement were annotated to the LabChart recording in real time by a researcher watching the live video of the spittlebug inside the chamber. Once a spittlebug had secreted a volume of froth that completely covered its body, the chamber was sharply tapped several times with a pair of metal forceps to simulate a potential predator. Every individual that made a spittle mass experienced 2–5 separate tapping events, and any changes in their CO2 following these ‘attacks’ were recorded.

Density determination

To determine the body density of both nymphs and adults, individuals were weighed in air and then while submerged in water, using an electronic balance (XPE205 balance and density determination kit, Mettler Toledo). Before being weighed underwater, individuals were placed in a syringe filled with soapy water. The plunger was pushed in and out repeatedly to completely flood the insect's tracheal system. Nymphs and adults were then transferred individually to a submerged weighing pan and re-weighed underwater. Body density (ρspittlebug) was calculated as:
formula
(4)
where Ma is the mass in air, Msub is the apparent mass underwater, ρ0 is the density of water (1 g cm−3) and ρL is the density of air (0.0012 g cm–3). These density measurements were used to correct the volume of air in the constant-volume respirometer by subtracting the volume of the insect from that of the vial.

Visualisation of O2 consumption within spittlebug foam

A novel fluorescence approach was used to visualise whether spittle bugs were capable of extracting O2 from the air bubbles trapped within the foam, as well as whether diffusion of O2 between adjacent air bubbles could occur. To do this, glass microscope slides (75×25×1 mm) were coated in a thin silicone elastomer film containing an O2-sensitive fluorophore: platinum (II) meso-tetra(pentafluorphenyl)porphine (PtTFPP; Frontier Scientific, Newark, DE, USA). The PtTFPP was first dissolved in toluene (concentration 1 mg ml−1), sonicated for 10 s, then added to the base component of a 2-part PDMS silicone (Silgard 184, Dow-Corning, Midland, MI, USA), for a final concentration of 0.05 mass per cent (the ratio of the mass of the PtTFPP to the mass of the total mixture, multiplied by 100). The toluene was then evaporated off using a vacuum centrifuge (Eppendorf Vacufuge Concentrator, Eppendorf Canada, Mississauga, ON, Canada). The PtTFPP and PDMS base was combined with the PDMS catalyst in a ratio of 10:1 base:catalyst and thoroughly mixed. This mixture was placed under a vacuum to remove any air bubbles incorporated during mixing, then a droplet was placed in the middle of a clean microscope slide. The droplet was left to spread over the surface of the slide, before the slide was placed on a temperature-controlled Peltier plate set to 40°C overnight. The final thickness of the PDMS film was 0.09 mm, measured using digital Vernier callipers.

To measure any change in PO2 within a spittle mass containing a nymph, a square frame with internal dimensions of 17×17 mm, wall thickness of 3 mm and height of 0.8 mm was 3D printed from ABS plastic. This was laid flat in the middle of the PDMS film. A nymph was placed into the square frame and its spittle mass was deposited on top of it, submerging it completely. A coverslip was then placed over the frame, sandwiching the foam and spittlebug inside the frame. A 1 mm gap cut into the middle of one of the square's sides allowed excess foam to escape from beneath the coverslip. The frame and coverslip held the spittlebug securely against the PDMS film within the foam, while preventing it from being crushed by the weight of the coverslip. The microscope slide was then placed on the stage of an inverted epifluorescence light microscope (IX73, Olympus, Tokyo, Japan). Excitation light (390 nm peak) from an X-Cite 120 LED light source (Excelitas Canada Inc., Vaudreuil-Dorion, QC, Canada) was used to excite the PtTFPP, while the emitted light was passed through a 645 nm (75 nm bandwidth) emission filter (ET645/75 m, Chroma Technology Corp., Bellows Falls, VT, USA) before being imaged using an Orca-Flash 4.0 LT CMOS camera (Hamamatsu, Hamamatsu City, Japan) using CellSens imaging software (Olympus, Tokyo, Japan). Images were taken at 10 min intervals. The fluorescence intensity of the PDMS film over time was measured at five locations in each image: the air bubble associated with the tip of the spittlebug's abdomen and four other, non-adjacent, air bubbles. ImageJ was used to calculate the average fluorescence intensity within each of these areas. Increases in fluorescence intensity within each bubble, associated with decreases in PO2, were quantified by dividing the fluorescence intensity within each bubble at each time point by the initial fluorescence intensity within the bubble at time zero.

Data analysis

All O2 and CO2 values used in the calculation of RQ were averaged from the last 1.75 h of the respirometry trials to reduce any influence of active movement on the measurements. RQs were calculated by dividing the mean, mass-specific CO2 by the mean, mass-specific O2. The CO2 of nymphs feeding on xylem was a grand mean averaged from all active foam production periods, and CO2 for walking was the grand mean of all continuous walking periods during the 2 h trials. All mass, O2 and CO2 values were log-transformed before linear regressions were applied to them.

Means, standard deviations, linear regressions and Johnson–Neyman tests were calculated in Microsoft Excel, and all other statistical tests and graphs were made using R (version 3.3.2; www.r-project.org) inside RStudio (version 1.1.383).

Field measurements of PO2 within established masses of spittlebug foam

There was no relationship found between PO2 and depth within the masses of foam measured in the field (F1,304=0.5368, P=0.4643) (Fig. 2). The mean (±s.d.) PO2 within the spittle was 19.97±1.17 kPa, and did not vary from the edge to centre of the spittle (n=306). This PO2 was not significantly different to the atmospheric level measured directly outside the spittle mass.

Fig. 2.

PO2 transects within masses of spittle measured in the field.PO2 was measured at regular 1 mm intervals from the outer edge (0 mm) to the centre of the mass. The PO2 values recorded at each interval (a total of 306 measurements) have been spread out horizontally to better show their distribution. As not all spittle masses (n=70) were the same size, there are fewer data points further from the edge. The solid line shows the linear regression.

Fig. 2.

PO2 transects within masses of spittle measured in the field.PO2 was measured at regular 1 mm intervals from the outer edge (0 mm) to the centre of the mass. The PO2 values recorded at each interval (a total of 306 measurements) have been spread out horizontally to better show their distribution. As not all spittle masses (n=70) were the same size, there are fewer data points further from the edge. The solid line shows the linear regression.

Laboratory measurements of PO2 within newly secreted foam

For all newly secreted masses of spittle, the PO2 was very slightly lower inside the foam than in the air outside it. This difference was statistically significant at the beginning of the trial (paired t-test, t13=6.65, P=1.59e−05). At the beginning of the 2 h measurement period, the mean (±s.d.) PO2 inside the freshly produced spittle was 20.1±0.42 kPa (n=14), and continued to increase to an average of 20.94±0.39 kPa (n=14) by the end of the 2 h.

Bubble chamber PO2

The PO2 within bubble chambers formed by moulting final instar spittlebugs was always lower than that in the outside air (paired t-test, t22=9.4581, P=3.294e−09) by an average of 1.91±0.97 kPa (n=23). It was observed that the PO2 would slowly rise during the measurement in association with the gradual drying out of the wet foam covering the bubble chamber. When the bubble ultimately popped, the PO2 immediately increased to the ambient level.

Density determination

In order to calculate the O2 consumption of individuals, the body density of both spittlebug nymphs and adults was determined. Philaenusspumarius nymphs have an average density of 1.11±0.0145 g cm−3 (n=3) and adults have an average density of 1.05±0.02 g cm−3 (n=3).

Metabolic rates and RQ

After log transformation, the nymphs' CO2 (µl h−1) followed the equation log(CO2)=0.845log(M)+0.0725, and O2 (µl h−1) followed the equation log(O2)=0.7649log(M)+0.1872, where M is nymph mass in mg. Using the Johnson–Neyman technique, spittlebug nymph CO2 and O2 while at rest were not found to be statistically different among slopes (F1,85=0.2124, P=0.6461) but were found to be statistically different among elevations (F1,86=5.55, P=0.0207; Fig. 3). The resting CO2 and O2 of adults, however, did not statistically differ among slopes (F1,37=0.581, P=0.451) or elevations (F1,38=0.467, P=0.5; Fig. 4). The CO2 of adult froghoppers followed the equation log(CO2)=1.39log(M)−0.5, and O2 followed the equation log(O2)=1.095log(M)−0.209. Using these measured CO2 and O2 rates, the RQ for P. spumarius was calculated to be 0.92 for nymphs and 0.95 for adults.

Fig. 3.

Resting metabolic rate (RMR) of P. spumarius nymphs at 20°C as a function of their mass. Mass was measured in mg; open circles are O2 measurements (μl h−1; n=43); filled circles are CO2 measurements (μl h−1; n=46). The dashed line and solid line are linear regressions for the O2 and CO2 data, respectively.

Fig. 3.

Resting metabolic rate (RMR) of P. spumarius nymphs at 20°C as a function of their mass. Mass was measured in mg; open circles are O2 measurements (μl h−1; n=43); filled circles are CO2 measurements (μl h−1; n=46). The dashed line and solid line are linear regressions for the O2 and CO2 data, respectively.

Fig. 4.

RMR of P. spumarius adults at 20°C as a function of their mass. Mass was measured in mg; open circles are O2 measurements (μl h−1; n=23); filled circles are CO2 measurements (μl h−1; n=18). The dashed line and solid line are linear regressions for the O2 and CO2 data, respectively.

Fig. 4.

RMR of P. spumarius adults at 20°C as a function of their mass. Mass was measured in mg; open circles are O2 measurements (μl h−1; n=23); filled circles are CO2 measurements (μl h−1; n=18). The dashed line and solid line are linear regressions for the O2 and CO2 data, respectively.

The CO2 of spittlebugs during locomotion followed the equation log(CO2)=0.464log(M)+0.478, whereas the CO2 of spittlebugs while actively feeding on xylem and excreting spittle followed the equation log(CO2)=0.342log(M)+0.6695. The CO2 of spittlebugs while making spittle was compared with production while walking using the Johnson–Neyman technique and they were found to follow the same slope (F1,26=0.11, P=0.743) but had different elevations (F1,27=5.26, P=0.03). CO2 during spittle production and feeding was on average 20±0.1% higher than when the same individual was walking (n=10, 3 outliers removed; Fig. 5). These values were compared with the RMR values, which showed that the slopes for RMR and metabolic rate during xylem feeding did not differ (F1,30=4.078, P=0.054), but their elevations did differ (F1,31=20.41, P<0.001; Fig. 5). RMR did not differ in slope (F1,26=2.82, P=0.1) or elevation (F1,27=1.679, P=0.21) from walking metabolic rate.

Fig. 5.

Metabolic rate of nymphs actively feeding on xylem or walking. Mass was measured in mg; filled circles are CO2 of nymphs feeding on xylem (μl h−1; n=17); open circles are CO2 of walking  nymphs (μl h−1; n=13). Lines indicate linear regressions for feeding (solid), walking (dashed) and calculated RMR (dotted).

Fig. 5.

Metabolic rate of nymphs actively feeding on xylem or walking. Mass was measured in mg; filled circles are CO2 of nymphs feeding on xylem (μl h−1; n=17); open circles are CO2 of walking  nymphs (μl h−1; n=13). Lines indicate linear regressions for feeding (solid), walking (dashed) and calculated RMR (dotted).

After a spittlebug had made a foam mass that completely covered it, the respirometry chamber was tapped sharply with metal forceps. Immediately after each knock, the nymph was observed to withdraw completely into the spittle. This movement was accompanied by an immediate drop in CO2 close to the chamber's baseline before returning back to the average production value that the nymph had had during feeding and spittle production. This drop in CO2 lasted an average of 46.5±7.84 s (n=17; Fig. 6). There was no correlation between the length of time CO2 dropped and the mass of the spittlebug nymph (F1,15=0.557, P=0.47).

Fig. 6.

Duration of nymph submergence in their spittle following mechanical disturbance. The duration of a spittlebug's submergence in spittle was determined from the length of time their CO2 decreased after tapping the respirometry chamber containing a nymph in a mass of spittle. Values are means from the 2–5 knocks each individual received (n=17). Error bars represent ±s.d.

Fig. 6.

Duration of nymph submergence in their spittle following mechanical disturbance. The duration of a spittlebug's submergence in spittle was determined from the length of time their CO2 decreased after tapping the respirometry chamber containing a nymph in a mass of spittle. Values are means from the 2–5 knocks each individual received (n=17). Error bars represent ±s.d.

Visualisation of O2 consumption within spittlebug foam

The first response of the spittlebugs forcibly submerged in the foam was to consolidate multiple smaller bubbles in the foam into a single large bubble (Fig. 7). This occurred as the mobile tip of the spittlebug's abdomen moved from bubble to bubble, breaking the surface tension between them, combining them into a single volume. For the duration of the measurement, the tip of the spittlebug’s abdomen remained within this air bubble. Over time, the level of fluorescence increased in the PDMS film in contact with this large bubble, but remained constant in all other regions of the foam (Fig. 8). A rise in fluorescence is consistent with a decreased PO2 in the large air bubble, and a concomitant decrease in O2-quenched fluorescence.

Fig. 7.

Decrease in PO2 in a single air bubble after a spittlebug was submerged within a mass of spittle. The spittlebug (ventral view) was forcibly submerged for 125 min in a spittle mass on a microscope slide coated in PDMS silicone film containing an O2-sensitive fluorophore (PtTFPP). The bubble associated with the tip of the spittlebug's abdomen (black arrow) and sternal groove containing the abdominal spiracles (white arrow) shows increased fluorescence, indicating a drop in PO2. Adjacent bubbles show no increase in fluorescence, indicating O2 levels remained stable.

Fig. 7.

Decrease in PO2 in a single air bubble after a spittlebug was submerged within a mass of spittle. The spittlebug (ventral view) was forcibly submerged for 125 min in a spittle mass on a microscope slide coated in PDMS silicone film containing an O2-sensitive fluorophore (PtTFPP). The bubble associated with the tip of the spittlebug's abdomen (black arrow) and sternal groove containing the abdominal spiracles (white arrow) shows increased fluorescence, indicating a drop in PO2. Adjacent bubbles show no increase in fluorescence, indicating O2 levels remained stable.

Fig. 8.

O2-dependent change in fluorescence intensity in five bubbles surrounding a spittlebug forcibly submerged in spittle. The tip of a nymph's abdomen was in the ‘breathing’ bubble, while the remaining four bubbles were not in contact with the tip of the spittlebug's abdomen. An increase in the fluorescence intensity (Ft)/initial fluorescence intensity (F0) ratio indicates a decrease in PO2 within the bubble. Lines are linear regressions.

Fig. 8.

O2-dependent change in fluorescence intensity in five bubbles surrounding a spittlebug forcibly submerged in spittle. The tip of a nymph's abdomen was in the ‘breathing’ bubble, while the remaining four bubbles were not in contact with the tip of the spittlebug's abdomen. An increase in the fluorescence intensity (Ft)/initial fluorescence intensity (F0) ratio indicates a decrease in PO2 within the bubble. Lines are linear regressions.

The absence of a sub-atmospheric PO2 within spittlebug foam, as well as the continuous CO2 measured from feeding spittlebugs using flow-through respirometry, shows definitively that P. spumarius snorkel: they breathe by extending the tip of their abdomen out of the foam mass, thereby bringing atmospheric air into contact with their ventral grove and abdominal spiracles. When a spittlebug retracts the tip of its abdomen into its foam mass, the measured CO2 drops immediately. Newly produced foam has a PO2 that is very slightly below atmospheric levels (<1 kPa), presumably as the incorporated air has just been expelled from the spittlebug's ventral groove. But this PO2 gradually rises over time to ambient levels, indicating that it is not subsequently being used for respiration. Even when startled, the insects would only submerge completely within their foam for 46.5 s, on average, and never longer than 70 s. This suggests that they do not attempt to access O2 from the bubbles of air trapped within the foam mass to prolong their submergence. Taken together, this evidence disproves the theory that spittlebugs use their foam as a ‘respiratory organ’ (Turner, 1994). That a wet foam is likely to offer a significant barrier to the uptake of atmospheric O2, despite the presence of air bubbles, has also been demonstrated in the aerial foam nests of frogs (Seymour and Loveridge, 1994). This same result can be seen clearly in Figs 7 and 8. After 125 min, only the air bubble that is in contact with the spittlebug's ventral groove shows any decrease in PO2, as indicated by the increase in fluorescence. There is no substantial diffusion of O2 from either the air bubbles or fluid in the surrounding foam. Thus, if an individual were to become trapped within its own spittle, it could use the tip of its abdomen to respire the O2 within the foam's air bubbles, but this appears to be an action of last resort.

PO2 within the bubble chamber

The one period when P. spumarius nymphs must remain submerged within their spittle occurs during the moult from final instar into adult. During this transition, the spittlebug produces a large single bubble of air in the centre of the foam (Fig. 1C). How they produce this chamber has not been determined, but it is entirely plausible that the spittlebug uses the same bubble-consolidation process as was observed when spittlebugs were forcibly submerged in their foam: the tip of their abdomen can break the surface tension of the spittle film around the bubbles, causing them to merge. Repeating this process within the centre of the spittle mass would consolidate the many small bubbles into a single large bubble chamber. However, the adult froghopper that emerges from the exuviae of the final instar must remain sealed within this bubble chamber for some period of time, as it possesses neither the flexible abdomen nor the ventral groove needed to break through the foam and access atmospheric O2. As a result, the PO2 inside the bubble chamber was on average 1.9 kPa lower than that of the ambient atmosphere during the first 15 min of measurement. Interestingly, all bubble chambers showed a gradual increase in PO2 during the 2 h measurement. This can be attributed to a gradual increase in the conductance of the spittle mass. Following moulting, the adult froghopper no longer feeds or adds excreted xylem fluid to the spittle mass. As a result, the spittle dries out, reducing the thickness of the liquid – the principal barrier to O2 diffusion through the foam. By the time the adult froghopper is ready to emerge, the foam usually has an almost powdery consistency and is probably quite permeable. From these measurements it appears that final instar spittlebugs and newly moulted adults are unlikely to ever experience more than a very minor hypoxia, even inside a completely enclosed bubble chamber.

Metabolic rate

From the resting O2 and CO2 values measured in this study, it can be seen that the nymphs and adults of P. spumarius have a RQ of 0.92 and 0.95, respectively, a finding that is entirely reasonable for an insect that feeds exclusively on the sugars and amino acids in plant sap. The resting O2 of spittlebug nymphs measured in this study also agrees closely with previous measurements on P. spumarius nymphs. Wiegert (1964) used a manometric respirometry technique and found that nymph O2 was on average 9.2 μl h−1 for fifth instars at 25°C, compared with the 9.8 μl h−1 calculated for an 11 mg (fifth instar) nymph in this study.

Metabolic cost of feeding on xylem

Feeding on xylem sap is assumed to be energetically expensive because of the presence of negative pressures within the xylem (Kim, 2013; Novotny and Wilson, 1997; Raven, 1983). Despite the expected cost of this feeding strategy, previous studies on P. spumarius have concluded that this species is capable of rapidly extracting large volumes of xylem sap under tensions exceeding 10 bar (1 MPa), as measured using a Scholander pressure bomb (Malone et al., 1999). Other studies have also found that the high fluid excretion rates of xylem-feeding sharpshooters (Hordinia circellata) are unaffected by manipulations that should increase the hydrostatic tension within the xylem of their host plant by 1 bar (Mittler, 1967). However, feeding rates in an adult leaf hopper (Homalodisca coagulata) decreased dramatically once xylem tension exceeded 1 MPa (Andersen et al., 1992). These findings are curious and suggest two possible explanations: either these bugs are capable of routinely producing pressures in excess of −1 MPa with their cibarial pumps, regardless of the metabolic cost, or the xylem pressures they experience while feeding are far lower than measured using a pressure bomb (Kim, 2013). On mechanical grounds it has been argued that a cibarial pump should not be able to produce a suction much greater than 0.3 MPa (Raven, 1983), supporting the latter possibility.

By measuring the metabolic rate of a spittle bug as it feeds on xylem, however, we have quantified the energetic cost of this activity; feeding P. spumarius spittlebugs had metabolic rates that were on average only 20% higher than their RMR. In a 9 mg insect, this is equivalent to a net feeding cost of ∼1.82 µl O2 h−1, or 0.0378 J h−1 (assuming a RQ of 0.92 and an oxyjoule equivalent of 20.75 J ml−1 O2; Lighton, 2008). The rate of fluid excretion during these metabolic rate measurements was also high, with the spittlebug being completely covered by a foam mass within 23.22±2.8 min (n=9). This suggests that the cost of extracting xylem sap is low.

A low feeding cost associated with xylem extraction is also reasonable considering the very low sugar content of xylem sap. Assuming that xylem sap contains 0.3 g glucose l−1 (1.66 mol m−3 of reducing sugars; Raven, 1983) and the spittlebug had an excretion rate of 72 μl xylem h−1 (20 nl s−1; Malone et al., 1999), then a feeding spittlebug should be able to obtain a maximum of 21.6 μg sugar h−1 if 100% of the sugar was extracted. Based on the insect's total metabolic rate of 0.23 J h−1, then extracting this quantity of xylem would provide approximately 1.5 times the insect's total metabolic rate requirements.

Furthermore, from the net increase in metabolic rate of a feeding spittlebug, it is possible to estimate the cibarial pump pressure. If a feeding spittlebug has a net feeding cost (xylem pumping cost) of 0.0378 J h−1 (1.05×10−5 J s−1 or W) and a volumetric flow of xylem of 20 nl s−1 (2×10−11 m3 s−1), then the cibarial pump pressure would be approximately −0.53 MPa. However, this calculation neglects the less than 100% efficiency of muscle. Assuming that muscle efficiency in insects lies between 10% and 50%, depending on the muscle type [skeletal muscle 50% (Pybus and Tregear, 1973), flight muscle 10% (Ellington, 1985)], this would reduce the cibarial pump pressure proportionately, to −0.26 to −0.05 MPa. For a spittlebug to draw up xylem sap, the pressure produced by their cibarial pump must be must be more negative than that of the sap in the xylem vessel. Pump pressure must also be sufficiently low to overcome the pressure drop that is attributable to the resistance of the spittlebug's stylets (which may or may not be substantial). This means that the xylem pressure of the sheep sorrel plants in these experiments must be less negative than the pump pressures calculated here. This would place the calculated pressures towards the lower end of the range reported from maize xylem (−0.7 to −0.1 MPa), as measured using both the Scholander pressure bomb and implanted pressure probe techniques (Wei et al., 1999). They are also below the theoretical maximum cibarial pump pressure of −0.3 MPa as estimated by Raven (1983). Thus, it remains to be seen whether the low pressures calculated here are representative of the xylem pressures experienced by feeding spittlebugs, or whether these insects are indeed capable of feeding from xylem with tensions exceeding 1 MPa, as has been suggested (Malone et al., 1999).

Besides P. spumarius, the most comprehensive analysis on the cost of fluid feeding in insects has been undertaken using Rhodnius prolixus, a triatomid hemipteran that sucks blood from a vertebrate host, also using a cibarial pump. So, how does the cost of xylem feeding compare with the cost of blood feeding? Fluid feeding was found to increase the metabolic rate of R. prolixus by up to 17-fold its pre-feeding metabolic rate (Leis et al., 2016). This dramatic increase in metabolic rate led Leis et al. (2016) to conclude that haematophagy is metabolically very expensive. But this conclusion ignores the fact that blood-feeding insects, and in particular R. prolixus, significantly depress their metabolic rate in-between meals (Bradley et al., 2003). Thus, given a low enough pre-feeding RMR, the factorial increase to even a routine metabolic rate (by the standards of other insects) will appear exceedingly high. Comparing the absolute mass-specific metabolic rate before and during feeding for both P. spumarius and R. prolixus reveals this difference. Pre-feeding R. prolixus have a metabolic rate of 0.00296 J h−1 mg−1 (Leis et al., 2016) compared with that of P. spumarius of 0.021 J h−1 mg−1 (calculated from the O2 of a 9 mg nymph using a RQ of 0.92 and the same oxyjoule equivalent of 20.75 J ml−1 O2). Thus, a pre-feeding spittlebug has a mass-specific metabolic rate that is 7.1 times higher than that of R. prolixus. In comparison, R. prolixus feeding on saline solution (as a reasonable proxy for xylem sap, but without the negative tension) has a mass-specific metabolic rate of 0.028 J h−1 mg−1 (Leis et al., 2016) compared with 0.025 J h−1 mg−1 in the xylem-feeding P. spumarius. This reveals that the mass-specific metabolic rates of these two bugs during fluid feeding are very similar, with the spittlebug's metabolic rate actually being slightly lower. However, differences in cibarial pump morphology, feeding rates, stylet resistance, fluid pressure and viscosity would all factor into interpreting these costs. For example, average pumping frequency in R. prolixus is 6 Hz (Smith, 1979), compared with the average of 1.3 Hz for a spittlebug measured in this study. The volume of fluid per pump is also much higher in R. prolixus [57.8 nl pump−1 in R. prolixus (Smith, 1979) versus 11.8 nl pump−1 in P. spumarius (Malone et al., 1999)]. Thus, understanding the differences in the form and function of cibarial pumps has the potential to reveal much about the limitations of fluid feeding in these remarkable insects.

Conclusions

This study shows that spittlebug nymphs breathe atmospheric O2 continuously by extending the tip of their abdomen out of the spittle mass. This allows these insects to continuously take up O2 and release CO2 while feeding on xylem. The increase in metabolic rate that accompanied xylem feeding is modest compared with their metabolic rate during rest and locomotion. Estimations of xylem pressure from the cost of feeding appear reasonable, and certainly within the range of xylem tensions measured in other studies. However, further studies should be done to investigate the metabolic rate, and rates of xylem excretion, of spittlebugs feeding on known xylem tensions to better understand the true metabolic costs associated with extracting this sap from plants.

We thank the two reviewers for their constructive criticism of this paper. Thanks also to Hsin-Yun Tsai and Russ Algar for their assistance with the vacuum centrifuge, and Karen Needham for species identification.

Author contributions

Conceptualization: P.G.D.M.; Methodology: K.I.S.B., A.B.R., P.G.D.M.; Formal analysis: A.B.R.; Investigation: K.I.S.B.; Writing - original draft: K.I.S.B., P.G.D.M.; Writing - review & editing: K.I.S.B., A.B.R., P.G.D.M.; Visualization: K.I.S.B., A.B.R., P.G.D.M.; Supervision: P.G.D.M.; Funding acquisition: P.G.D.M.

Funding

This work was supported by the Natural Sciences and Engineering Research Council of Canada [Undergraduate Student Research Award to K.I.S.B., CSG-M to A.B.R., Discovery grant: RGPIN-2014-05794 to P.G.D.M.].

Andersen
,
P. C.
,
Brodbeck
,
B. V.
and
Mizell
,
R. F.
(
1992
).
Feeding by the leafhopper, Homalodisca coagulata, in relation to xylem fluid chemistry and tension
.
J. Insect Physiol.
38
,
611
-
622
.
Bradley
,
T. J.
,
Brethorst
,
L.
,
Robinson
,
S.
and
Hetz
,
S.
(
2003
).
Changes in the rate of CO2 release following feeding in the insect Rhodnius prolixus
.
Physiol. Biochem. Zool.
76
,
302
-
309
.
del Campo
,
M. L.
,
King
,
J. T.
and
Gronquist
,
M. R.
(
2011
).
Defensive and chemical characterization of the froth produced by the cercopid Aphrophora cribrata
.
Chemoecology
21
,
1
-
8
.
Ellington
,
C. P.
(
1985
).
Power and efficiency of insect flight-muscle
.
J. Exp. Biol.
115
,
293
-
304
.
Guilbeau
,
B. H.
(
1908
).
The origin and formation of the froth in spittle-insects
.
Am. Nat.
42
,
783
-
798
.
Hamilton
,
K. G. A.
(
1982
).
The spittlebugs of Canada, Homoptera: Cercopidae
. In
The Insects and Arachnids of Canada
, part 10.
Ottawa
:
Agriculture Canada, Research Branch
.
Horsfield
,
D.
(
1978
).
Evidence for xylem feeding by Philaenus spumarius (L.) (Homoptera:Cercopidae)
.
Entomol. Exp. Appl.
24
,
95
-
99
.
Kim
,
W.
(
2013
).
Mechanics of xylem sap drinking
.
Biomed. Eng. Letters
3
,
144
-
148
.
Kosztarab
,
M.
(
1979
).
Observations on the Spittlebug, Tomaspis insignita Fowler (Homoptera: Cercopidae)
.
Fla. Entomol.
62
,
396
-
401
.
Leis
,
M.
,
Pereira
,
M. H.
,
Casas
,
J.
,
Menu
,
F.
and
Lazzari
,
C. R.
(
2016
).
Haematophagy is costly: respiratory patterns and metabolism during feeding in Rhodnius prolixus
.
J. Exp. Biol.
219
,
1820
-
1826
.
Li
,
Q.
,
Zhong
,
H.
,
Zhang
,
Y.
and
Wei
,
C.
(
2015
).
Comparative morphology of the distal segments of Malpighian tubules in cicadas and spittlebugs, with reference to their functions and evolutionary indications to Cicadomorpha (Hemiptera: Auchenorrhyncha)
.
Zool. Anz.
258
,
54
-
68
.
Lighton
,
J. R. B.
(
2008
).
Measuring Metabolic Rates: a Manual for Scientists
.
New York
;
Oxford
:
Oxford University Press
.
Malone
,
M.
,
Watson
,
R.
and
Pritchard
,
J.
(
1999
).
The spittlebug Philaenus spumarius feeds from mature xylem at the full hydraulic tension of the transpiration stream
.
New Phytol.
143
,
261
-
271
.
Marshall
,
A. T.
(
1966
).
Spittle-production and tube-building by cercopid larvae (Homoptera)—IV. Mucopolysaccharide associated with spittle-production
.
J. Insect Physiol.
12
,
635
-
644
.
Marshall
,
A. T.
(
1973
).
Protein synthesis and secretion by the Malpighian tubules of cercopoid larvae (Homoptera)
.
J. Insect Physiol.
19
,
2317
-
2326
.
Mello
,
M. L. S.
,
Pimentel
,
E. R.
,
Yamada
,
A. T.
and
Storopoli-Neto
,
A.
(
1987
).
Composition and structure of the froth of the spittlebug, Deois SP
.
Insect Biochem.
17
,
493
-
502
.
Mittler
,
T. E.
(
1967
).
Water tensions in plants—an entomological approach
.
Ann. Entomol. Soc. Am.
60
,
1074
-
1076
.
Novotny
,
V.
and
Wilson
,
M. R.
(
1997
).
Why are there no small species among xylem-sucking insects?
Evol. Ecol.
11
,
419
.
Pybus
,
J.
and
Tregear
,
R.
(
1973
).
Estimates of force and time of actomyosin interaction in an active muscle and of the number interacting at any one time
.
Cold Spring Harbor Symp. Quant. Biol.
37
,
655
-
660
.
Rakitov
,
R. A.
(
2002
).
Structure and function of the Malpighian tubules, and related behaviors in juvenile cicadas: Evidence of homology with spittlebugs (Hemiptera: Cicadoidea & Cercopoidea)
.
Zool. Anz.
241
,
117
-
130
.
Raven
,
J. A.
(
1983
).
Phytophages of xylem and phloem: a comparison of animal and plant sap-feeders
. In
Advances in Ecological Research
,
Vol. 13
, pp.
135
-
234
.
Elsevier
.
Seymour
,
R. S.
and
Loveridge
,
J. P.
(
1994
).
Embryonic and larval respiration in the arboreal foam nests of the African frog Chiromantis xerampelina
.
J. Exp. Biol.
197
,
31
-
46
.
Smith
,
J. B. J.
(
1979
).
The feeding response of Rhodnius prolixus to blood fractions, and the role of ATP
.
J. Exp. Biol.
78
,
225
-
232
.
Tonelli
,
M.
,
Gomes
,
G.
,
Silva
,
W. D.
,
Magri
,
N. T. C.
,
Vieira
,
D. M.
,
Aguiar
,
C. L.
and
Bento
,
J. M. S.
(
2018
).
Spittlebugs produce foam as a thermoregulatory adaptation
.
Sci. Rep.
8
,
1
-
6
.
Turner
,
J. S.
(
1994
).
Anomalous water loss rates from spittle nests of spittlebugs Aphrophora saratoga (Homoptera: Cercopidae)
.
Comp. Biochem. Physiol.
107
,
679
-
683
.
Weaver
,
C. R.
and
King
,
D. R.
(
1954
).
Meadow spittlebug, Philaenus leucophthalmus (L.)
.
Wooster, OH, USA
:
Ohio Agricultural Experiment Station
.
Wei
,
C.
,
Tyree
,
M. T.
and
Steudle
,
E.
(
1999
).
Direct measurement of xylem pressure in leaves of intact maize plants. a test of the cohesion-tension theory taking hydraulic architecture into consideration
.
Plant Physiol.
121
,
1191
-
1205
.
Whittaker
,
J. B.
(
1970
).
Cercopid Spittle as a microhabitat
.
Oikos
21
,
59
-
64
.
Wiegert
,
R. G.
(
1964
).
Population energetics of meadow spittlebugs (Philaenus spumarius l.) as affected by migration and habitat
.
Ecol. Monogr.
34
,
218
-
241
.

Competing interests

The authors declare no competing or financial interests.