ABSTRACT
We present a generic sub-diffraction-limited imaging method – photobleaching imprinting microscopy (PIM) – for biological fluorescence imaging. A lateral resolution of 110 nm was measured, more than a twofold improvement over the optical diffraction limit. Unlike other super-resolution imaging techniques, PIM does not require complicated illumination modules or specific fluorescent dyes. PIM is expected to facilitate the conversion of super-resolution imaging into a routine lab tool, making it accessible to a much broader biological research community. Moreover, we show that PIM can increase the image contrast of biological tissue, effectively extending the fundamental depth limit of multi-photon fluorescence microscopy.
INTRODUCTION
Fluorescence microscopy has been extensively used to gain a deeper understanding of cell and tissue biology (Lichtman and Conchello, 2005). Despite widespread applications, fluorescence microscopy faces a fundamental bottleneck in its resolving capability, the optical diffraction limit, which was first stated by Ernst Abbe in 1873. Owing to this limitation, the achievable spatial resolution of conventional fluorescence microscopy is ∼250 nm in the lateral direction and ∼500 nm in the axial direction. Because many cellular organelles, such as microtubules, actin fibers and ribosomes, are smaller than this, breaking the optical diffraction limit has been the holy grail of light microscopy over the past several decades.
To address this challenge, a wide array of methodologies has been introduced, allowing the fine structure of a biological cell or tissue to be revealed at the super-resolution level. Depending on the conceptual shared thread, super-resolution techniques are normally based on two distinct strategies. One strategy, referred to as probe-based super-resolution, uses the stochastic nature of photoswitchable fluorophores, sequentially switching them on and off so that the signals from individual fluorescent molecules can be recorded consecutively. Merging all of the single-molecule centroid positions acquired by repeated cycles of photo-activation and photo-deactivation yields a final super-resolution image. Representative modalities using this strategy are stochastic optical reconstruction microscopy (STORM) (Rust et al., 2006), photoactivated localization microscopy (PALM) (Betzig et al., 2006; Hess et al., 2006) and photobleaching-based localization microscopy (Burnette et al., 2011; Munck et al., 2012; Simonson et al., 2011). The second strategy, referred to as illumination-based super-resolution, uses non-linear optical approaches to modify and engineer the point spread function (PSF), resulting in a reduced focal spot size at the sample. Within this category, the most important techniques are stimulated emission depletion microscopy (STED) (Klar et al., 2000), structured illumination microscopy (SIM) (Gustafsson, 2000) and saturated structured illumination microscopy (SSIM) (Gustafsson, 2005). Owing to its superior resolving capability, super-resolution imaging now has become an essential tool for examining a wide variety of biological molecules, pathways and dynamics, providing unprecedented resolution at a scale down to tens of nanometers (Hell, 2007).
In spite of their revolutionary impact, most super-resolution techniques rely on specific fluorescent probes, such as PALM/STORM, or require complicated optical illumination modules, such as STED, limiting access by the general research community. To provide a generic method that can be readily implemented on a standard light microscope with conventional fluorescent dyes, here we present photobleaching imprinting microscopy (PIM) for super-resolution fluorescence imaging. Photobleaching is routinely encountered as a nuisance in fluorescence microscopy (Tsien et al., 2006), which is leveraged here. PIM works by first imprinting a structured pattern onto the sample through photobleaching, followed by light interrogation with another focused Gaussian laser beam. The high-order components related to the laser fluence distribution can be extracted simply by a differential operation between two fluorescence intensities along time-lapse fluorescence decay. Using PIM, we demonstrated a lateral resolution of ∼110 nm, more than a twofold improvement over the optical diffraction limit.
PIM can also remove the out-of-focus light, enhancing image contrast and enabling deep imaging. In two-photon fluorescence microscopy (TPFM), the in-focus signals are ultimately limited by near-surface fluorescence (Theer and Denk, 2006). Here we show that, at a depth of 300 µm, whereas TPFM showed almost no contrast between the in-focus signals and background, PIM clearly resolved the sample features. Therefore, PIM is expected to effectively extend the fundamental imaging depth limit of optical microscopy for characterization of deep tissues.
Operating principle
To show a potential resolution improvement in theory, we simulated a case when a fluorophore, aminocoumarin dextran, was imaged by confocal fluorescence microscopy (CFM) and PIM under two-photon excitation. The original object, the CFM image and the PIM image are shown in Fig. 1D–F, respectively. The simulation algorithm is detailed in Materials and Methods.
RESULTS AND DISCUSSION
Sub-diffraction-limited fluorescence PIM
We first evaluated the lateral resolution of PIM by imaging a sharp edge of a thin fluorescent film (maximal emission wavelength ∼520 nm) with a confocal fluorescence microscope. With an oil-immersion objective with a numerical aperture (NA) of 1.4, the full width at half maximum (FWHM) of the PSF measured by PIM was ∼110 nm, less than half of the ∼280 nm measured by conventional confocal microscopy (Fig. 2A).
Then we imaged a thin layer of aggregated fluorescent microspheres (diameter ∼0.9 µm, maximum emission wavelength ∼600 nm) sandwiched between two cover glasses. The fluorescent microspheres were excited and photobleached by a He-Ne laser at 543 nm. The pixel dwell time was 200 µseconds, and a total of 200 time-lapse frames were captured. The PIM image intensities were calculated pixel by pixel by a differential operation between two consecutive images and averaged along the fluorescence decay curve (see Materials and Methods). As shown in Fig. 2B, as a result of the crosstalk from adjacent pixels, the contrast in conventional confocal fluorescence images was low. In the corresponding PIM image, owing to the improvement in resolution, the fluorescent microspheres became more distinguishable.
To demonstrate the sub-diffraction-limited imaging capability of PIM in a biological cell, we imaged cellular microtubules (bovine pulmonary artery endothelial cell line) stained with a green fluorescent dye, BODIPY FL Goat Anti-Mouse IgG (F-14781, Life Technologies). The fluorophore was excited and photobleached by an argon laser at 488 nm. Following a similar procedure, 200 time-lapse fluorescence images were measured to extrapolate the PIM image, shown in Fig. 2C. The signal intensities across a line in the zoomed area were plotted and compared between the confocal fluorescence and PIM images (Fig. 2D). Note that at the locations near 4 µm and 6 µm, the microtubules are not discriminable in the confocal fluorescence image, whereas they are well resolved in the corresponding PIM image. In addition, we also imaged cellular mitochondria (NIH 3T3 fibroblast cell line) stained with an orange fluorophore, Mitotracker CMT (M7510, Life Technologies). The sample was excited and photobleached by a He-Ne laser at 543 nm. The acquired confocal fluorescence and PIM images are shown in Fig. 2E. To highlight the difference, a side-by-side image is shown in Fig. 2F. Because PIM relies on fluorescence decay to extract the nonlinear components, we evaluated the gain in resolution improvement against the loss of fluorescence intensity level after the PIM procedure. A measurement showed that the cellular fluorescent intensity was decreased by ∼10% owing to photobleaching, indicating the feasibility of multi-cycle PIM imaging before the fluorophore loses its entire fluorescing capability.
Contrast enhancement in multi-color tissue fluorescence imaging by PIM
In CFM, sectioning is achieved by using a pinhole to block the out-of-focus light. However, when the pinhole size is sufficiently small (in practice, smaller than 0.25 Airy unit), the thickness of the optical section is no longer influenced by the pinhole size but solely determined by the axial resolution of the microscope objective in use. By improving the resolution in all dimensions, PIM can reduce the crosstalk from adjacent voxels, thus allowing super-contrast imaging at depths.
To demonstrate the contrast enhancement capability of PIM in tissue fluorescence imaging, we imaged a mouse kidney section (∼20 µm thick) stained with a combination of three fluorophores (F24630, Life Technologies). First, the elements of the glomeruli and convoluted tubules were labeled with green fluorescent Alexa Fluor 488. Second, the filamentous actin prevalent in glomeruli and the brush border were stained with red fluorescent Alexa Fluor 568. Third, the nuclei were counterstained with a blue-fluorescent dye, DAPI. The fluorophores were excited at wavelengths of 405 nm, 488 nm and 543 nm. The emitted fluorescence was collected by a microscope objective (NA = 1.4) and simultaneously measured in three separate photomultiplier tube (PMT) color channels. An optimal pinhole size was automatically chosen by the microscope to balance the in-focus SNR and optical sectioning. For PIM calculation, a total of 400 time-lapse fluorescence images were captured.
To study the contributions to the PIM image of different order nonlinear components in Eq. 6, we fitted a polynomial function to the measured fluorescent decay pixel by pixel (see Materials and Methods). The images associated with the linear, quadratic and cubic terms of the fluence distribution were extracted and are shown in Fig. 3A, respectively. The green and red fluorescent intensities along the dashed line in the zoomed area were plotted and compared between the linear and nonlinear components (Fig. 3B–E). As expected, a higher-order dependence on the laser fluence led to a higher image contrast.
Extending the fundamental imaging depth limit in two-photon fluorescence microscopy by PIM
Two-photon fluorescence microscopy (TPFM) is an indispensable technique for deep tissue imaging (Helmchen and Denk, 2005; Piston, 1999). As a result of a quadratic dependence on the laser fluence, the excitation in TPFM is spatially confined, enabling intrinsic optical sectioning without a physical pinhole. However, this optical sectioning capability eventually fails at a deeper depth, because the out-of-focus fluorescence gradually overwhelms the in-focus signal in a scattering sample. The depth where the out-of-focus signals are comparable to the in-focus fluorescence is defined as the fundamental imaging depth limit of TPFM (Theer and Denk, 2006).
To demonstrate how the contrast enhancement capability of PIM extends the imaging depth limit in TPFM, we imaged a phantom consisting of aggregated green fluorescent microspheres covered by a 300 µm thick micro-fiber layer stained with fluorescein. The sample was excited by a Ti-sapphire femtosecond laser at 920 nm, and the two-photon excited fluorescence was collected by a water-immersion objective with an NA of 1.04. To provide a baseline, we first directly imaged the fluorescent microspheres without the scattering layer depicted in Fig. 4A to acquire the reference image shown in Fig. 4B. Then, we covered the sample with the scattering layer (Fig. 4C) and imaged over the same FOV from the top (Fig. 4D). Owing to the fluorescence from the out-of-focus layers, the sample features indicated with an arrow in Fig. 4B are completely obscured in Fig. 4D. Moreover, the fluorescent intensities across the dashed line in Fig. 4D show no contrast between the in-focus signal and background (Fig. 4F). Next, we acquired a PIM image by measuring 400 time-lapse fluorescence images (Fig. 4E). The signal intensities across the same line are shown in Fig. 4F, demonstrating the recovery of the indicated features.
Conclusion
In summary, we present a generic sub-diffraction-limited technique, photobleaching imprinting microscopy, for biological fluorescence imaging. In phantom and cellular imaging experiments, PIM demonstrated over a twofold improvement in resolution. Additionally, we showed that PIM can reduce the out-of-focus light in tissue fluorescence imaging, considerably improving the image contrast. Compared with other super-resolution imaging techniques, PIM is easier to implement on a standard microscope with generic fluorescent dyes, making super-resolution imaging a routine lab technique. It is worth noting that, although techniques such as deconvolution microscopy can also remove image blur, the accuracy is highly dependent on the signal-to-background ratio (Murray, 2011). In particular, the extension of fundamental depth in TPFM as seen in Fig. 4 would not even be possible by solely relying on deconvolution microscopy.
Although not demonstrated here, PIM could also be potentially used in wide-field fluorescence microscopy and light-sheet fluorescence microscopy (LSFM). Owing to a structured fluence distribution along the depth axis, on the one hand, wide-field fluorescence PIM can be used to remove the out-of-focus light, thereby allowing for optical sectioning; on the other hand, light-sheet fluorescence PIM can be exploited to further improve axial resolution, offering a simple solution to a long-standing problem in LSFM on mitigating the trade-off between imaging field-of-view and section thickness (Fahrbach et al., 2013).
The resolution improvement by PIM is achieved at the expense of fluorescence loss. A 10% drop in fluorescence intensity was observed after the PIM procedures. In addition, averaging the differential intensity along the fluorescence decay course is normally required to obtain a sufficient SNR to extrapolate the high-order components, a process that decreases the temporal resolution of the microscope and currently allows monitoring of only slow-varying events. This limitation can be potentially overcome by choosing a smaller FOV, scanning a path of interest with random access, or shortening the pixel dwell time using a higher fluorophore concentration. Moreover, parallel acquisition schemes, such as spinning-disk confocal microscopy, can also be used. Because PIM is based on the photobleaching memory effect, the achievable resolution is dependent on the light fluence distribution on the focal plane. In deep imaging, the focal diameter deviates from the diffraction limit because of light scattering (Liu et al., 2012), a fact that also degrades the resolution of PIM. Consequently, in practice, the imaging depth of PIM is still restricted by the optical diffusion limit, which is ∼1 mm in biological tissue.
For convenient operation, the data acquisition in the presented work was conducted in a frame-based manner. At each sampling pixel, the time interval between two consecutive excitation events is limited by the frame acquisition time, which can be relatively long owing to a large number of spatial sampling pixels. This limitation might compromise the PIM image formation in live cultures because phenomena such as fluorescence recovery after photobleaching (FRAP) can erase the memory of a sample on the previous light field distribution. To overcome this problem, the data acquisition of PIM can be switched to a point-based approach, measuring the entire fluorescence decay at a sampling pixel before the scanner moves to the next location. This method can minimize the time interval between two consecutive excitation events down to the integration time of the detector (200 µseconds in the presented study), a time scale within which the diffusion of fluorophores can be generally neglected.
MATERIALS AND METHODS
Numerical simulation for confocal and PIM fluorescence imaging
Fluorescence microscopy
The PIM experiments were implemented on a confocal laser-scanning microscope equipped with a multi-photon imaging unit (FV1000, Olympus). Two continuous-wave laser sources (Argon and He-Ne) provide a total of four wavelengths (405 nm, 488 nm, 543 nm and 632 nm) for one-photon excitation, and a Ti-sapphire femtosecond pulsed laser provides a tuneable wavelength (780–1060 nm) for two-photon excitation. The fluorescence was collected by a high NA microscope objective (PLAPON 60×O, NA = 1.4, for one-photon excitation; 25× MPE water immersion, NA = 1.05, for two-photon excitation) and was simultaneously detected by up to three separate PMT color channels. The laser excitation and acquisition of fluorescence images were controlled by the accompanying microscope software (Fluoview, Olympus).
Photobleaching imprinting microscopy (PIM)
PIM is achieved by monitoring photobleaching-induced fluorescence decay, followed by a pixel-by-pixel differential operation between two consecutive fluorescent images. The PIM procedure can be performed by using the time-lapse imaging module on a standard confocal microscope. To make surrounding regions unaffected at each sampling pixel, the scanning step size must be larger than the diffraction-limited PSF during acquiring a time-lapse dataset. Then we translate the sample a distance that is equal to half of the resolution of PIM and start the next cycle. The PIM images are calculated for each cycle and stitched together to form a final image.
Cell culture and fluorescent staining
NIH 3T3 fibroblast cells were cultured in Dulbecco's modified Eagle's medium supplemented with 10% fetal bovine serum, 2 mM glutamine, and 1% penicillin-streptomycin. The cells were incubated at 37°C in 5% CO2 and divided every ∼72 hours. To prepare the cell slides, the fibroblast cells were seeded to a cover glass surface. The culture medium was removed 24 hours after seeding and replaced by the staining solution, 2 µM MitoTracker Orange probes (M-7510, Life Technologies) in fresh culture medium. After incubation in staining solution for 60 minutes, the cells were rinsed with fresh medium and then with phosphate buffered saline (PBS). The cell slides were then fixed in 4% formaldehyde for 15 minutes and then rinsed with PBS before imaging.
Acknowledgements
We thank J. Ballard for close reading of the manuscript. We would also like to thank L. D. Wang and J. Yao for helpful discussions.
Author contributions
L.G. designed the biological experiment, performed part of the microscopy experiments, analyzed the data, and prepared the manuscript. A.G. analyzed part of data, and prepared the manuscript. Y.L. contributed to experimental design and performed part of microscopy experiments. C.L. prepared the biological sample and performed part of the microscopy experiments. L.V.W. contributed to the conceptual system and experimental design and manuscript preparation.
Funding
This work was sponsored in part by National Institutes of Health (NIH) [grant numbers DP1 EB016986 (NIH Director's Pioneer Award), R01 EB008085, R01 CA134539, U54 CA136398, R01 CA157277 and R01 CA159959]. Deposited in PMC for release after 12 months.
References
Competing interests
L.W. has a financial interest in Microphotoacoustics, Inc. and Endra, Inc., which, however, did not support this work.